CALIBRATE 210 Plus
- Complete visual Inspection of pipette for functional integrity
- Extensive servicing, cleaning and preventive maintenance performed, inside as well as outside
- Piston assembly servicing: polishing / corrosion removal / Lubrication
- Replace seal and o-ring kit, check the sealing system (free for select brands)
- Replace broken or worn-out parts (at additional cost)
- Check pipette functionality–physical alignment and mechanical
- Perform leak test on sealing system
- Re-Calibrate using “Gravimetric Methodology”
- Precision and accuracy verification with 10 weighs (readings) each at low point and high point through the range of the pipette.
- Labeling (ID/Serial number / QC date / Due date / Status of Calibration)
- Document performance Verification and generate certificate of calibration (Optional – available upon request prior to service, no additional charge)
Back